Specificity Calculator
Interactive companion to the Battery Method Validation technical note. Enter your ICP-OES / ICP-MS specificity data and receive instant numerical results with interpretation.
LOD Criterion
Blank signals must be below the Limit of Detection
Inputs
Results
LOD Calculation
σ_blank = 60 cps
LOD = 3 × 60 = 180 cps
Blank Assessment
Reagent blank = 45 cps 45 < 180 ✓
Matrix blank = 62 cps 62 < 180 ✓
Polyatomic Interference Check
No-gas vs. He-KED mode comparison (ICP-MS)
Inputs
Results
Ratio Calculation
No-gas = 48,200 cps
He-KED = 2,310 cps
Ratio = 48,200 / 2,310 = 20.9×
Interpretation
Polyatomic contribution = 95.2% of no-gas signal
→ He-KED / DRC mode required ✗
Matrix Effect
Sensitivity comparison — aqueous vs. matrix-matched
Inputs
Results
Matrix Effect Calculation
Saq = 2,310 cps/(µg/L)
Smatrix = 2,199 cps/(µg/L)
Effect = (2,199 − 2,310) / 2,310 × 100
= -4.8%
|-4.8%| ≤ 10% ✓
Internal Standard Recovery
Verify IS signal is within the acceptance window
Inputs
Results
IS Recovery Calculation
Expected = 125,000 cps
Measured = 119,375 cps
Recovery = (119,375 / 125,000) × 100
= 95.5%
85% ≤ 95.5% ≤ 115% ✓