METHOD VALIDATION · SPECIFICITY

Specificity Calculator

Interactive companion to the Battery Method Validation technical note. Enter your ICP-OES / ICP-MS specificity data and receive instant numerical results with interpretation.

1

LOD Criterion

Blank signals must be below the Limit of Detection

Inputs

Results

LOD Calculation

σ_blank = 60 cps

LOD = 3 × 60 = 180 cps

Blank Assessment

Reagent blank = 45 cps  45 < 180  ✓

Matrix blank  = 62 cps  62 < 180  ✓

Both blank signals are below the LOD. No detectable signal contribution from the reagent or matrix — specificity criterion satisfied.
2

Polyatomic Interference Check

No-gas vs. He-KED mode comparison (ICP-MS)

Inputs

Results

Ratio Calculation

No-gas = 48,200 cps

He-KED = 2,310 cps

Ratio = 48,200 / 2,310 = 20.9×

Interpretation

Polyatomic contribution = 95.2% of no-gas signal

→ He-KED / DRC mode required ✗

Significant polyatomic interference (95.2% of no-gas signal). He-KED or DRC mode is required. Standard acquisition would report a severely biased result.
3

Matrix Effect

Sensitivity comparison — aqueous vs. matrix-matched

Inputs

Results

Matrix Effect Calculation

Saq = 2,310 cps/(µg/L)

Smatrix = 2,199 cps/(µg/L)

Effect = (2,199 − 2,310) / 2,310 × 100

= -4.8%

|-4.8%| ≤ 10%  ✓

Matrix effect within the ±10% acceptance limit. Aqueous calibration is valid; internal standard correction is sufficient.
4

Internal Standard Recovery

Verify IS signal is within the acceptance window

Inputs

Results

IS Recovery Calculation

Expected = 125,000 cps

Measured  = 119,375 cps

Recovery = (119,375 / 125,000) × 100

= 95.5%

85% ≤ 95.5% ≤ 115%  ✓

IS recovery of 95.5% is within the 85–115% acceptance window. Internal standard correction is performing as expected.