Determine the validated operating window by confirming linearity, accuracy, and precision
simultaneously at each boundary concentration. Range is accepted only when all three
parameters pass at every endpoint.
Calibration Curve — used for linearity check at each boundary
Slope (S)
signal / conc
Intercept (b)
signal units
ŷ = S × C + b · residual% = (measured − ŷ) / ŷ × 100
Boundary 1ALL PASS ✓
Concentration
LinearityPASS ✓
Measured Signal
Max |Residual| (%)
ŷ = 0.00185
residual = 1.62%
accept: ±2.0%
AccuracyPASS ✓
Min Rec%
Max Rec%
RSD max
Spike Recoveries (%)
mean = 98.70%
RSD = 0.46%
accept: 80–120%, RSD ≤ 10%
PrecisionPASS ✓
RSD max (%)
Replicate Values
mean = 0.5000
SD = 0.0155
RSD = 3.10%
accept: RSD ≤ 10%
Boundary 2ALL PASS ✓
Concentration
LinearityPASS ✓
Measured Signal
Max |Residual| (%)
ŷ = 0.19100
residual = 0.30%
accept: ±2.0%
AccuracyPASS ✓
Min Rec%
Max Rec%
RSD max
Spike Recoveries (%)
mean = 100.10%
RSD = 0.26%
accept: 90–110%, RSD ≤ 5%
PrecisionPASS ✓
RSD max (%)
Replicate Values
mean = 20.0000
SD = 0.1404
RSD = 0.70%
accept: RSD ≤ 5%
Validated Range
Boundary
Conc.
Linearity
Accuracy
Precision
Overall
Lower — 0.5 ppm (LOQ)
0.5
✓1.62%
✓98.70%
✓3.10%
PASS
Upper — 20.0 ppm (200% spec)
20
✓0.30%
✓100.10%
✓0.70%
PASS
Range Visualization
✓ Range Accepted
0.5 to 20
Linearity, accuracy, and precision confirmed at all 2 boundary points.