GATE 06

Range Calculator

Determine the validated operating window by confirming linearity, accuracy, and precision simultaneously at each boundary concentration. Range is accepted only when all three parameters pass at every endpoint.

Calibration Curve — used for linearity check at each boundary
Slope (S)
signal / conc
Intercept (b)
signal units

ŷ = S × C + b  ·  residual% = (measured − ŷ) / ŷ × 100

Boundary 1 ALL PASS ✓
Concentration
Linearity PASS ✓
Measured Signal
Max |Residual| (%)
ŷ = 0.00185
residual = 1.62%
accept: ±2.0%
Accuracy PASS ✓
Min Rec%
Max Rec%
RSD max
Spike Recoveries (%)
mean = 98.70%
RSD = 0.46%
accept: 80–120%, RSD ≤ 10%
Precision PASS ✓
RSD max (%)
Replicate Values
mean = 0.5000
SD = 0.0155
RSD = 3.10%
accept: RSD ≤ 10%
Boundary 2 ALL PASS ✓
Concentration
Linearity PASS ✓
Measured Signal
Max |Residual| (%)
ŷ = 0.19100
residual = 0.30%
accept: ±2.0%
Accuracy PASS ✓
Min Rec%
Max Rec%
RSD max
Spike Recoveries (%)
mean = 100.10%
RSD = 0.26%
accept: 90–110%, RSD ≤ 5%
Precision PASS ✓
RSD max (%)
Replicate Values
mean = 20.0000
SD = 0.1404
RSD = 0.70%
accept: RSD ≤ 5%
Validated Range
BoundaryConc.LinearityAccuracyPrecisionOverall
Lower — 0.5 ppm (LOQ)0.5 1.62% 98.70% 3.10%PASS
Upper — 20.0 ppm (200% spec)20 0.30% 100.10% 0.70%PASS
Range Visualization
VALIDATED RANGE0.520
✓ Range Accepted
0.5 to 20
Linearity, accuracy, and precision confirmed at all 2 boundary points.