METHOD VALIDATION · ACCURACY

Accuracy Calculator

Enter spike recovery measurements across concentration levels. The calculator computes % recovery, SD, and RSD per level with overall summary — matching the Battery Method Validation walkthrough.

Analyte & Units

C_spiked = ppm
PASS

Acceptance Criteria

Measured Values (ppm)

n = 3
#C_found
1
2
3

Individual Recoveries

#C_found (ppm)Recovery %
10.48096.0%
20.510102.0%
30.49098.0%

n = 3

Mean = 98.67%  (80–120%)  ✓

SD = 3.06%

RSD = 3.10%  (≤ 10%)  ✓

C_spiked = ppm
PASS

Acceptance Criteria

Measured Values (ppm)

n = 3
#C_found
1
2
3

Individual Recoveries

#C_found (ppm)Recovery %
19.94099.4%
210.080100.8%
310.120101.2%

n = 3

Mean = 100.47%  (90–110%)  ✓

SD = 0.95%

RSD = 0.94%  (≤ 5%)  ✓

C_spiked = ppm
PASS

Acceptance Criteria

Measured Values (ppm)

n = 3
#C_found
1
2
3

Individual Recoveries

#C_found (ppm)Recovery %
119.87099.4%
220.140100.7%
320.060100.3%

n = 3

Mean = 100.12%  (90–110%)  ✓

SD = 0.69%

RSD = 0.69%  (≤ 5%)  ✓

Overall Summary — Fe (m/z 56)

PASS
Total Measurements
n = 9
Grand Mean Recovery
99.75%
Overall RSD
1.84%

Overall SD = 1.833%

Levels evaluated: 3 / 3

All spike levels pass recovery and RSD criteria. Grand mean = 99.75%, overall RSD = 1.84%. Method demonstrates adequate accuracy for Fe (m/z 56).