Accuracy Calculator
Enter spike recovery measurements across concentration levels. The calculator computes % recovery, SD, and RSD per level with overall summary — matching the Battery Method Validation walkthrough.
Analyte & Units
Acceptance Criteria
Measured Values (ppm)
n = 3Individual Recoveries
| # | C_found (ppm) | Recovery % |
|---|---|---|
| 1 | 0.480 | 96.0% |
| 2 | 0.510 | 102.0% |
| 3 | 0.490 | 98.0% |
n = 3
Mean = 98.67% (80–120%) ✓
SD = 3.06%
RSD = 3.10% (≤ 10%) ✓
Acceptance Criteria
Measured Values (ppm)
n = 3Individual Recoveries
| # | C_found (ppm) | Recovery % |
|---|---|---|
| 1 | 9.940 | 99.4% |
| 2 | 10.080 | 100.8% |
| 3 | 10.120 | 101.2% |
n = 3
Mean = 100.47% (90–110%) ✓
SD = 0.95%
RSD = 0.94% (≤ 5%) ✓
Acceptance Criteria
Measured Values (ppm)
n = 3Individual Recoveries
| # | C_found (ppm) | Recovery % |
|---|---|---|
| 1 | 19.870 | 99.4% |
| 2 | 20.140 | 100.7% |
| 3 | 20.060 | 100.3% |
n = 3
Mean = 100.12% (90–110%) ✓
SD = 0.69%
RSD = 0.69% (≤ 5%) ✓
Overall Summary — Fe (m/z 56)
PASSOverall SD = 1.833%
Levels evaluated: 3 / 3